II.C.2 Cardiovascular monitoring
Arterial and central lines
PA catheter and TEE
ECG and hemodynamic monitoring
What each cardiovascular monitor actually measures, how it lies to you, and how to prove the number on the screen is real.
By the end you can
- Set up ECG leads, filters and a blood pressure cuff so the displayed numbers can be trusted
- Level, zero and dynamically test an arterial line, and read its waveform
- Interpret CVP, wedge and cardiac output data, and name what invalidates each one
Your case
OR 12 · Open aortic aneurysm repair · 40 min in
Mr. R, 74, 82 kg. Coronary disease, hypertension. Right radial arterial line, right internal jugular central line, 5-lead ECG.
The arterial trace looks rounded. The dicrotic notch is gone. Meanwhile the cuff on the other arm reads 124/72.
Art line 96/78 (86)
Cuff 124/72 (88)
CVP 14
HR 96
Hold that thought
- The line and the cuff systolics disagree by almost 30 mmHg. Which one is closer to the truth, and how do you prove it?
- His CVP is 14. Does that mean he has all the volume he needs?
- You ask for a cleaner ECG trace and the ST segment changes. What just happened?
You will answer these at the end of the lesson.
Members only
The rest of this lesson is for members
Membership opens all 98 lessons, every cheat sheet and memory card, the full question bank, and simulation exams.
Not ready to join? Free account: take the 50-question diagnostic, see your weakest domain, and get a study plan built around your exam date. No card.
Start free
Inside this lesson
- Start
- The case
- Three leads or five
- Filters and artifact
- The cuff
- Picking a site
- Level and zero
- Reading the waveform
- Damping tester
- Central lines
- The CVP waveform
- Pulmonary artery catheter
- Output monitors and TEE
- Lock it in
- Board traps
- Cheat card
- 5-question quiz
- Wrap-up