II.A.3 Flowmeters
II.A.4 Vaporizers
I.D.2 Gas laws and vaporization
Flowmeters and vaporizers
The two dials you touch most, and the physics that decides whether the patient gets what you set.
By the end you can
- Explain why a flowmeter tube is tapered, gas specific, and placed with oxygen downstream
- Calculate what a variable bypass vaporizer actually delivers from the splitting ratio
- Predict output when temperature, back pressure, altitude, or the wrong agent changes
Your case
Surgery center at 5,280 feet · shoulder scope · 20 min in
Healthy 42 year old, general anesthesia with a laryngeal mask. Total fresh gas flow 2 L/min. Your sevoflurane dial reads 2.0%.
The agent analyzer reads 2.5% sevoflurane. The CRNA who had the room before you says the vaporizer has "been reading high all week."
Dial 2.0%
Analyzer 2.5%
FGF 2 L/min
BP 108/62
Hold that thought
- Is the vaporizer broken, or is 2.5% exactly what you should expect here?
- The next room runs desflurane on a Tec 6. Does that dial need a different number?
- Later you drop to 0.4 L/min and start positive pressure ventilation, and the analyzer creeps up again. What is that?
You will answer these at the end of the lesson.
Members only
The rest of this lesson is for members
Membership opens all 98 lessons, every cheat sheet and memory card, the full question bank, and simulation exams.
Not ready to join? Free account: take the 50-question diagnostic, see your weakest domain, and get a study plan built around your exam date. No card.
Start free
Inside this lesson
- Start
- The case
- Inside the flowmeter
- Why it is tapered
- Reading it right
- Sequence and leaks
- Inside a vaporizer
- Staying accurate
- Back pressure
- Tipping and misfilling
- Desflurane and altitude
- Modern machines
- Vaporizer output explorer
- Lock it in
- Board traps
- Cheat card
- 5-question quiz
- Wrap-up