II.A.1 Medical gas supply
II.A.2 Cylinders and pipelines
II.A.5 to II.A.7 Machine safety
Gas supply: pipelines, cylinders, failsafes
Where your oxygen comes from, how long the tank on the back will last, and what to do the day the wall lies to you.
By the end you can
- Convert a cylinder gauge reading into liters and into minutes of use
- Name every safety system between the wall and the patient, and what each one misses
- Run a pipeline failure and a pipeline crossover in the right order
Your case
OR 6 · Lap sigmoid colectomy · 2 hours in
Mrs. K, 62, stable all morning. You are running 1 L/min oxygen and 1 L/min air with sevoflurane.
The low inspired oxygen alarm fires. You look up:
Everything else on the machine looks normal. Nothing else is alarming.
SpO2 88%
O2 analyzer 17%
Pipeline gauge 52 psig
EtCO2 36
Hold that thought
- Why did only one machine monitor catch this?
- What do you do, and in what order?
- Your E cylinder reads 1,200 psig. At 6 L/min, how long do you have?
You will answer these at the end of the lesson.
Members only
The rest of this lesson is for members
Membership opens all 98 lessons, every cheat sheet and memory card, the full question bank, and simulation exams.
Not ready to join? Free account: take the 50-question diagnostic, see your weakest domain, and get a study plan built around your exam date. No card.
Start free
Inside this lesson
- Start
- The case
- Where oxygen comes from
- E cylinders
- The nitrous gauge
- Colors and connections
- Three pressure systems
- Regulators and valves
- The fail-safe
- Hypoxic guard
- Cylinder bench
- Run the numbers
- Crossover
- Lock it in
- Board traps
- Cheat card
- 5-question quiz
- Wrap-up